Maintenance manual one White Flower Shiitake Mushroom Dride White Flower Shiitake Mushroom,Flower Shiitake Mushroom,Fresh Shiitake,Fresh Shiitake Mushrooms Near Me SHANDONG JOIN & SHARE AGRICULTURAL DEVELOPMENT CO., LTD , https://www.joinsharemushroom.com
1. The use of the instrument must be carried out by a special person. When the equipment is used, the external power supply should be cut off to avoid accidents.
2. When placing the sample in the working room, it should not be too dense, so as not to affect the normal flow of the airflow in the working room. There must be a gap of 10-20mm between each item, and it should be placed symmetrically. The items on each shelf should be staggered. The distance between the bottom shelf and the bottom of the studio must be greater than 100 mm.
3. A certain space should be left around the equipment to facilitate heat dissipation, operation and maintenance of the equipment.
4. Prevent sharp and sharp objects from scratching the device.
5. Do not disassemble the electrical components and other parts of the equipment at will, so as not to damage the electrical control circuit, causing human error of the equipment and affecting its performance.
6. When storing and using the equipment, there are no flammable, explosive and corrosive substances around.
7. Do not put flammable, explosive and volatile materials into the equipment for drying.
8. Please read the “Product User's Manual†carefully before using the equipment, and then perform various operations.
1) Before using, you must pay attention to whether the power supply voltage used is consistent. When using, the power outlet ground wire must be grounded as specified.
2) When using the power, do not touch the electrical parts of the space on the left side of the box with your hands or use a damp cloth and rinse with water. The power should be cut off during maintenance.
3) The power cord should not be wrapped around metal objects. It should not be placed in a place with high temperature or humidity to prevent the rubber from aging and causing leakage.
4) Do not over-extrude the contents of the box, and leave room for natural convection of the air so that the humid air can accelerate out on the wind top.
5) Regularly check whether the silver contact of the temperature regulator is hairy or uneven. If it is, use the fine gauze to level the contact, then use it, and always wipe it with a cleaning cloth to make it in good contact (be careful to cut off) power supply). Do not hit the metal pipe of the indoor temperature regulator to avoid affecting the sensitivity.
6) Do not put flammable materials in the dry box without explosion-proof device.
7) After each use, the power supply must be cut off, and the inside and outside of the box should be kept clean. Hang the corresponding logo after cleaning.
Maintenance manual II
1. Explosive, flammable and volatile materials should not be placed in a dry box to avoid explosion.
2. The average load of the sample shelf should not exceed 15kg.
3. No items can be placed on the shelf between the sample chamber and the heating chamber to avoid affecting heat exchange.
4, must check the presence or absence of overlap or short circuit of the nickel wire before use.
5, the heating indicator does not light, the general reasons are three: 1 bulb is broken; 2 bulb contact is bad; 3 heating system failure.
Maintenance and maintenance of electric drying oven:
6. Turn on the heating switch and do not heat it. There are generally four reasons: 1 nickel-chromium wire breaking; 2 switch and line desoldering; 3 temperature control knob refers to zero position (rotation temperature control knob to the specified temperature scale); 4 relay damage.
7, temperature control failure, the general reasons are three: 1 temperature probe damage (replace the thermistor); 2 relay damage; 3 temperature control system other components failure (using a universal meter to detect the fault, according to the situation).
Maintenance manual three
1. Before using, you must pay attention to whether the power supply voltage used is consistent. When using, the power outlet ground wire must be grounded as specified.
2. When using the power, do not touch the electrical parts of the space on the left side of the box with your hands or use a damp cloth and rinse with water. The power should be cut off during maintenance.
3. The power cord should not be wrapped around metal objects. It should not be placed in a place with high temperature or humidity to prevent the rubber from aging and causing leakage.
4, the contents of the box should not be over-squeezed, you must leave room for natural convection of air, so that humid air can accelerate on the wind top.
5. Regularly check whether the silver contact of the temperature regulator is hairy or uneven. If it is, use the fine gauze to level the contact and then use it, and always wipe it with a cleaning cloth to make it in good contact (be careful to cut off) power supply). Do not hit the metal pipe of the indoor temperature regulator to avoid affecting the sensitivity.
6. Do not put flammable materials in the dry box without explosion-proof device.
7. After each use, the power supply must be cut off, and the inside and outside of the box should be kept clean. Hang the corresponding logo after cleaning.